Hioki Leak Current HiTester ST5540

Hioki Leak Current HiTester ST5540

  • Uninterrupted polarity switching function
  • Safety conductor current measurement function
  • Automatic measurement functionality
  • 110% voltage application jack
  • Circuit breaker for device under measurement
  • Save measurement data for 100 devices
  • Improved test reliability
  • Simple, interactive operation
  • Ability to store up to 30 sets of measurement conditions

 

REQUEST A QUOTE
SKU: ST5540 Category:



Introducing the Hioki Leak Current HiTester ST5540

Complies with all standards (suitable for use with all networks)

Leak current parameters as defined for medical-use electrical devices include ground leak current, contact current, patient leak current, and patient measurement current. The ST5540 provides a single solution for measuring all of these leak current variants.

Complies with IEC 60601-1:2005 +A1:2012 (Ed 3.1)

In order to prevent the danger of electric shock, electrical devices use power supplies that are isolated from parts of the device that may come into contact with the body. However, it is impossible to achieve infinite insulation resistance. Some leak current always exists, and its magnitude changes as the insulation degrades over time. The LEAK CURRENT HiTESTER ST5540/ST5541 provides an easy-to-operate solution for measuring leak current in electrical devices, making it eminently suitable for use in an extensive array of applications, ranging from production lines to equipment maintenance and inspections.

Hioki Leak Current HiTester ST5540


Key Benefits

For medical-use electrical devices and essential to electrical safety

Compliance with IEC 60601-1:2005 Ed 3.0, JIS T 0601-1:2012

(*Starting on June 1, 2012, medical electrical equipment sold in the EU must comply). Model ST5540 comply with IEC 60601-1:2005 +A1:2012 (Ed 3.1), and IEC 62353 of 2017

Compliance with Electrical Appliances and Materials Safety Act,

JIS, IEC, and UL standards for general-use electrical devices

• Uninterrupted polarity switching function dramatically reduces cycle time

• Support for rated currents up to 20 A gives the instrument more than adequate capability for testing products designed to comply with new standards

 


Hioki Leak Current HiTester ST5540


Trust the Experts at ACA TMetrix Inc.

ACA TMetrix Inc. is a leading Canadian distributor of test and measurement instruments and design tools. For over 55 years we have provided products manufactured by the world’s leading instrument manufacturers. Leading Distributor of Design Tools and Test Equipment in Canada.

Specifications

Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)

Measurement methods Measurement of voltage drop across body simulated resistance points, Calculation and display of current values, True rms measurement, Measurement unit floats relative to instrument ground.
Measurement modes Leak current measurement, voltage measurement, safety conductor current measurement
Standards compliance (NW: Body simulated resistance) [NW-A] • Electrical Appliances and Materials Safety Act
[NW-B1] • Medical electrical equipment: IEC 60601-1:1988+ A1:1993+ A2:1995, JIS T 0601-1:1999
[NW-B2] • Medical electrical equipment: IEC 60601-1:2005+ A1:2012, JIS T 0601-1:2012 and complement 1:2014, IEC 62353
[NW-C]
• Measurement of touch current and protective conductor current: IEC 60990:2016
• Electrical equipment for measurement, control, and laboratory use: IEC 61010-1:2010+ A1:2016
• Information technology equipment: IEC60950-1:2005+ A1:2009+ A2:2013
• Audio, video and similar electronic apparatus: IEC 60065:2014
• Personnel Protection Systems for EV: UL 2231-1:2012 (Amended 2016), UL-2231-2:2012 (Amended 2016)
[NW-D] • For UL: UL 1492:1996 (Amended 2013)
[NW-G] • Electrical equipment for measurement, control, and laboratory use; current measurement circuits in damp conditions: IEC 61010-1:2010+ A1:2016
Leak current measurement Ground leak current, 3 types of contact current, 7 types of patient leak current, patient measurement current, 4 types of total patient leak current, free current measurement, 3 types of enclosure leak current
Measurement current DC, AC (true rms, 0.1 Hz to 1 MHz), AC+DC (true rms, 0.1 Hz to 1 MHz), AC peak (15 Hz to 1 MHz)
Measurement ranges DC / AC / AC+DC mode: 50.00 mA/ 5.000 mA/ 500.0 μA/ 50.00 μA
AC peak mode: 75.0 mA/ 10.00 mA/ 1.000 mA/ 500.0 μA
Measurement accuracy (current measurement) DC measurement: ±2.0% rdg. ±6 dgt. (typ.)
AC / AC+DC measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 100 kHz, typ.)
AC peak measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 10 kHz, typ.)
Interfaces External I/O, medical device relay output, USB 1.1 (communications), RS-232C
Functionality 110% voltage application, automatic test, data storage for 100 target devices, clock, data backup, printed output (optional), etc.
Power supply 100/120/220/240 V AC (specify at time of order), 50/60 Hz, 30 VA rated power
Target device power supply input 100 to 250 V AC, 50/60 Hz Rated current input from terminal block: 20 A
Target device power supply output Output from terminal block: 20 A Output from outlet: 15 A
Dimensions and mass 320 mm (12.60 in)W × 110 mm (4.33 in)H × 253 mm (9.96 in)D, 4.5 kg (158.7 oz)
Accessories Test lead L2200 (for ST5540, Red ×2, Black ×1) ×1 set, Enclosure probe 9195 ×1, Power cord ×3, Spare fuse for measurement line ×1, Instruction manual ×1, User guide ×1, CD-ROM ×1

 