BATTERY HiTESTER BT3562

BATTERY HiTESTER BT3562

3 mΩ to 3000 Ω/60V DC Battery Testers for Production Line Testing of Battery Packs and Modules

REQUEST A QUOTE
SKU: BATTERY HiTESTER BT3562 Categories: , , ,



Description

Hioki benchtop battery testers support simultaneous high-speed measurement of internal resistance and battery voltage for the ever-expanding production lines of increasingly larger lithium-ion low resistance batteries, and other battery packs for high voltage applications. The BT3562 can measure the voltage and resistance of coin-size to large-cell batteries up to 60V almost simultaneously due to their quick response time. BATTERY HiTESTER BT3562

BT3562 basic model
BT3562-01 with GP-IB and analog output

Note: Measurement leads are not included. Purchase the appropriate lead option for your application separately. The male (system side) of the EXT I/O connector is also available. Please inquire with your Hioki distributor.

GPIBBT3562-01RS232CCE  

Key Features BATTERY HiTESTER BT3562

• Measure high-voltage battery packs up to 60V (BT3562)

• Production line testing of high-voltage battery packs and battery modules

• Large (low-resistance) cell testing

• Choice of PC interfaces for full remote operation Note: The comparison threshold values depend on the battery manufacturer, type, and capacity, and these must be established by the user.

■ Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)

 

Max. applied measurement voltage

 

± 60 VDC rated input voltage
± 70 VDC Max. rated voltage to earth

 

Resistance measurement ranges

 

3 mΩ (Max. display 3.1000 mΩ, resolution 0.1 mΩ) to 3000 Ω (Max. display 3100.0 Ω, resolution 100 mΩ), 7 ranges
Accuracy: 30 mΩ to 3000 Ω ranges, ± 0.5% rdg. ± 5 dgt. (Add ± 3 dgt. for EX.FAST, or ± 2 dgt. for FAST and MEDIUM)
3 mΩ range, ± 0.5% rdg. ± 10 dgt. (Add ± 30 dgt. for EX.FAST, or ± 10 dgt. for FAST, or ± 5 dgt. for MEDIUM)
Testing source frequency: 1 kHz ± 0.2 Hz, testing current: 100 mA (3 mΩ range) to 10 mA (3000 Ω range)
Open-circuit Voltage: 25 V peak (3/30 mΩ ranges), 7 V peak (300 mΩ range), 4 V peak (3 Ω to 3000 Ω ranges)

 

Voltage measurement ranges

 

6 VDC (resolution 10 mV) to 60 VDC (resolution 100 mV), 2 ranges

 

Accuracy: ± 0.01% rdg. ± 3 dgt. (Add ± 3 dgt. for EX.FAST, or ± 2 dgt. for FAST and MEDIUM)

 

Display

 

31000 full digits (resistance), 600000 full digits (voltage), LED

 

Sampling rate

 

Four steps, 4 ms (Extra-FAST), 12 ms (FAST), 35 ms (Medium), 150 ms (Slow)
(Typ., sampling time depends on supply frequency settings and function.)

 

Measurement time

 

Response time + sampling rate, approx. 10 ms for measurements
(Response time depends on reference values and the measurement object.)

 

Comparator functions

 

Judgment result: Hi/IN/Lo (resistance and voltage judged independently)
Setting: Upper and lower limit, Deviation (%) from reference value
Logical ANDed result: PASS/FAIL, calculates the logical AND of resistance and voltage judgment results.
Result display, beeper, or external I/O output, Open-collector (35 V, 50 mA DC max.)

 

Analog output

 

Measured resistance (displayed value, from 0 to 3.1 V DC, -01 suffix models only)

 

Interfaces

 

External I/O, RS-232C, Printer (RS-232C), GP-IB (-01 suffix models only)

 

Power supply

 

100 to 240 VAC, 50/60 Hz, 30 VA max.

 

Dimensions and mass

 

215 mm (8.46 in) W × 80 mm (3.15 in) H × 295 mm (11.61 in) D, 2.4 kg (84.7 oz)

 

Accessories

 

Instruction manual ×1, Power cord ×1