Specifications
General Specifications
Specifications: Valid after 15 minute warm-up, for 1 year from last external calibration, and for ambient temperature within +/- 2 Degrees C of last performed ZERO operation.
Ethernet: High speed, high noise immunity LAN interface
RS232 Interface: Selectable baud: 9600, 19200, 38400, 57600 or 115200, full handshake
VICL Interfaces: Provided for control of HV Scanners
Digital I/O Interface: Provides 4 digital inputs and 5 digital outputs. Functions include, Start/Stop, Testing, Pass/Fail, HV Present, Safety Interlock, Dwell Timer
Optional GPIB: Option GPIB-9 adds GPIB interface capability
Operating Temperature: 0°C to 50°C
Humidity: <85% RH (non-condensing)
Power: 105-265Vrms, 45-65Hz, 500VA Max
Dimensions: 89mm H x 432mm W x 457mm D (3.5″ H x 17″ W x 18″ D)
Weight: 5Kg (11 lb.) Net / 7Kg (15 lb.)
Accessories: Alligator test leads, NIST traceable calibration certificate with no data, power cord and operator’s manual. Limited ISO 17025 cal cert with data & uncertainties available.
DC Output Voltage
981i: 30V to 6500V
Accuracy: +/- (0.25% rdg + 0.1V)
981i with option HSS2: 30V to 5000V
Accuracy: +/- (0.25% rdg + 0.1V+1V/mA)
983i: 60V to 11000V
Accuracy: +/- (0.25% rdg + 0.2V)
Output Resolution: 1V
Loading: up to 4mA charging, 2mA continuous
Current Measurement
981i & 983i RETURN current only, 1pA to 4mA
Accuracy (Resistive Load up to 1000pF): +/- (0.5%rdg + 10pA + 2pA per kV))
Accuracy (Capacitive Load 1000pF to 50nF): +/- (0.5%rdg + 100pA + 20pA per kV))
Accuracy (Capacitive Load >50nF): +/- (0.5%rdg + 1nA + 200pA per kV))
Option HSS2 HV Output current, 1nA to 4mA (for grounded loads)
Accuracy (Resistive Load up to 1000pF), C<1000pF): +/- (0.5%rdg + 5nA + 1nA per kV)) Accuracy (Capacitive Load 1000pF to 50nF): +/- (0.5%rdg + 7.5nA + 2.5nA per kV)) Accuracy (Capacitive Load >50nF): +/- (0.5%rdg + 10nA + 5nA per kV))
Resistance Measurement
NOTE: the use of the OFFSET (or ZERO) capability in either AUTO or MANUAL TEST modes is not required to meet these specifications – the OFFSET capability is intended to allow the user to offset externally produced leakages.

Transfer Measurement
Comparative Measurement Capability (measurement of multiple values within 10% at the same test voltage, in the same environmental conditions within +/-1C, within a 1 hour period, with no intervening ZERO operations, and constant power)
Uncertainty (Resistive Load): +/- (0.05%rdg + (R/(5Tohm per kV))%rdg + (1/V)%rdg + 2ohms per kV) for R < 150Tohm
Timing
Charge Time: 1 second minimum charge time, 9999 sec max.
Dwell Time: 0.1s to 9999 days, 0.1 sec resolution