Description
Hioki LCR Meters and Impedance Analyzers range from 1mHz to 1.3GHz devices to suit a wide range of applications in the testing of electronic components. The IM3536 raises the industry standard for general-purpose LCR meters by offering a wide DC and 4Hz to 8MHz testing frequency while delivering 0.05% accuracy, ideal for evaluating the characteristics of next generation electronic components, including power supply inductors thanks to the instrument’s maximum measurement frequency of 8MHz. LCR METER IM3536
IM3536 |
---|
Note: Test fixtures are not supplied with the instrument. Select optional test fixtures or probes when ordering.
Key Features
• DC, 4Hz to 8MHz measurement frequency
• High-speed measurement of 1ms (fastest time)
• High-precision measurement of ±0.05% rdg. (representative value)
• Guaranteed accuracy range from 1 mΩ, low-impedance measurement with unmatched repeatability
• DC bias function: Measure under conditions simulating actual use or in accordance with industry standards
• Exceptional specifications and cost-performance for a wide range of applications, from R&D to production lines
■ Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year) LCR METER IM3536
Measurement modes | LCR mode, Continuous measurement mode |
---|---|
Measurement parameters |
Z, Y, θ, X, G, B, Q, Rdc (DC resistance), Rs (ESR), Rp, Ls, Lp, Cs, Cp, D (tanδ), σ, ε |
Measurement range |
100 mΩ to 100 MΩ, 10 ranges (All parameters are determined according to Z) |
Display range |
Z: 0.00 m to 9.99999 GΩ, Y: 0.000 n to 9.99999 GS, θ: ± (0.000° to 180.000°), Q: ± (0.00 to 9999.99), Rdc: ± (0.00 m to 9.99999 GΩ), |
Basic accuracy |
Z ±0.05% rdg. θ: ±0.03° (representative value, Measurable range: 1 mΩ to 200 MΩ) |
Measurement frequency |
4 Hz to 8 MHz (5 digits setting resolution, minimum resolution 10 mHz) |
Measurement signal level |
LCR METER IM3536
[Normal mode: V mode/CV mode] |
DC bias measurement |
Generating range: DC voltage 0 V to 2.50 V (10 mV resolution) |
Output impedance |
Normal mode: 100 Ω, Low impedance high accuracy mode: 10 Ω |
Display |
5.7-inch color TFT with touch panel |
Functions |
Comparator, BIN measurement (10 categories for 2 measurement parameters), Trigger function, Open/short compensation, Contact check, Panel loading/saving, Memory function |
Interfaces |
EXT. I/O( HANDLER) ,USB, USB flash drive, LAN, GP-IB, RS-232C, BCD |
Power supply |
100 to 240 V AC, 50/60 Hz, 50 VA max. |
Dimensions and mass |
330 mm (12.99 in) W × 119 mm (4.69 in) H × 230 mm (9.06 in) D, 4.2 kg (148.1 oz) |
Accessories |
Power cord ×1, Instruction manual ×1, LCR application disc (Communications user manual) ×1 |