Discount Products: Look through products available for a 25% – 50% discount in 2020. The items provided on a first come, first serve basis.
View Discounted Products

inTEST Thermal Temptronic ATS-605 ThermoStream®

inTEST Thermal Temptronic ATS-605 ThermoStream®

  • Customizable & savable test setups
  • Program & Datalog Storage
  • User Defined Temperature Limits
  • Local & Remote Operations
  • LabView™ drivers
  • IEEE-488, RS232 ports

 

 

REQUEST A QUOTE
SKU: ATS-605 Category:



Introducing the inTEST Thermal Temptronic ATS-605 ThermoStream®

Mobile Temperature Environments

Temptronic’s ThermoStreams are portable systems that deliver clean dry air for precision temperature testing or conditioning of electronics (ICs, MEMS, transceivers, or circuits) and materials. No other systems can bring your test subjects to temperature faster with precise control.

 

inTEST Thermal Temptronic ATS-605 ThermoStream®



inTEST Thermal Temptronic ATS-605 ThermoStream®

Key Features

FEATURES:

  • Customizable & savable test setups
  • Program & Datalog Storage
  • User Defined Temperature Limits
  • Local & Remote Operations
  • LabView™ drivers
  • IEEE-488, RS232 ports


Key Benefits

With rapid thermal cycling or long saturation at a precise temperature, these systems optimize throughput for characterization, verification, thermal shock or failure analysis applications – at the test bench or on the production floor.

  • No LN2 or LCO2 required
  • -100 to +300°C
  • Programmable transitions, up to 15°C/sec
  • 1.0°C accuracy
  • Extended reach for easy DUT access
  • Integrated purging system prevents moisture buildup on the DUT
  • NEW! Quick-connect feature for thermal caps

 


inTEST Thermal Temptronic ATS-605 ThermoStream®


Trust the Experts at ACA TMetrix Inc.

ACA TMetrix Inc. is a leading Canadian distributor of test and measurement instruments and design tools. For over 55 years we have provided products manufactured by the world’s leading instrument manufacturers. Leading Distributor of Design Tools and Test Equipment in Canada.

Specifications

Power 60Hz only, system does not operate at 50Hz
115 ±10% VAC, 60 Hz 15 amp, 1phase
230 ±10% VAC, 60 Hz 10 amp, 1phase
COMPRESSED AIR
Clean, Dry Air (CDA) Filtered to 5 micron particulate contamination.
Oil Content: <0.1 ppm, by weight, filtered to 0.01 micron oil contaminant.
Dewpoint: ambient to -40°C
Air Supply Pressure 5.6 to 7kg /cm2 (80 to 90 PSIG), 5.6 kg/cm (90 PSIG) nominal
Total Air Flow Rate Required 4.0 to 6.2 l/s (8-13 scfm), 4.5 l/s (9 scfm) nominal
Air Supply Temperature +20° to +28°C; +23°C nominal
OPERATING ENVIRONMENT
Operating Temperature +20° to +28°C; +23°C nominal
Humidity 0 to 60%; 45% nominal
WEIGHTS & DIMENSIONS
Dimensions
Controller: 40.1cm (15.8 inches) WIDE
14.8cm (5.8 inches) HIGH
48.9cm (19.2 inches) DEEP
Thermal Wand: 8cm (3 inches) DIAMETER
36cm (14 inches) LONG
Stand: 48.25cm (19 inches) approximately above
benchtop to top of post
Reach from Vertical Post: standard, 51cm (20 inches)
Weight Controller with Thermal Wand and Hose: 14.5 kg (32lbs.)
Thermal Wand Stand Assembly: 7kg (5.5 lbs.)
Thermal Cap: .45 kg (1lb.)
Noise Level <70dBA
SERVICE & SAFETY
Refrigerants
HCFC and CFC-free, non-toxic, non-flammable
Serviceability Auto-diagnostics and field replaceable modules
Over Temperature Protection +230°C (factory set): Operator can set high and low air temperature limits

 

 

Datasheet