PW9100-03 For the PW6001/3390, 3 ch
PW9100-04 For the PW6001/3390, 4 ch

 

REQUEST A QUOTE



Description

Hioki benchtop power meters and power analyzers are best in class power measuring instruments for measuring single to three-phase lines with a high degree of precision and accuracy. The PW9100 is a direct AC/DC current measurement option for the PW6001 and 3390 series Power Analyzers that delivers wide bandwidth and high accuracy testing across 3 channels of input. The option is ideal for applications requiring direct input connections when testing the power conversion efficiency of motors, primary and secondary side of inverters, hybrid and electric vehicles, and renewable energy equipment.  PW 9100 – 03

PW9100-03 For the PW6001/3390, 3 ch
PW9100-04 For the PW6001/3390, 4 ch
CE

Key Features  PW9100 – 03

• World-leading measurement bands and accuracy at a 50 A rating

• Wide-band DC to 3.5MHz, 50A AC/DC rated input

• ±0.04% power accuracy in combination with PW6001 ±0.04%*

• 120dB CMRR (100 kHz) • Full-rack size suitable for test/evaluation benches

• Current measurement option for PW6001/ 3390/ 3390-10 POWER ANALYZERS

      * For complete details, please refer to the specifications

■ Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)

Measurement line type Isolated input, DCCT input
Rated primary current 50 A AC/DC
Number of input channels PW9100-03: 3 channels
PW9100-04: 4 channels
Maximum input current 60 A, within derating. However, up to ±200 A peak is allowable if within 20 ms (design value)
Amplitude and Phase accuracy DC (±0.02 % rdg. ±0.007 % f.s.)
45 Hz < f ≤ 65 Hz (±0.02 % rdg. ±0.005 % f.s., Phase: ±0.1 deg.)
Accuracy is defined to 1 MHz
Output voltage 2 V/50 A
Measurement terminals Terminal block (with safety cover), M6 screws
Input resistance 1.5 mΩ or less (50 Hz/60 Hz)
Input capacitance Between measurement terminals and case (secondary side), 40 pF or less, defined at 100 kHz
Operating temperature and humidity Temperature: 0°C to 40°C (32°F to 104°F), Humidity: 80% RH or less (no condensation)
Power supply Power supply from PW6001, 3390, 3390-10
Dimensions and mass 430 mm (16.93 in) W × 88 mm (3.46 in) H × 260 mm (10.24 in) D, Cable length: 0.8 m (2.62 ft)
PW9100-03: 3.7 kg (130.5 oz), PW9100-04: 4.3 kg (151.7 oz)
Accessories Instruction Manual ×1