Temperature Performance1, 2 |
Range: -65 to 175°C @ 23°C ambient
Accuracy: ±1.0°C; Stability: ±0.3°C |
Cooling Power4 |
55W @ -55°C
125W @ -40°C
250W @ 0°C |
Transition Rates1 |
25 to -40°C <35 sec, 25 to +125°C <1.5 min. |
Temperature Sensors |
Main sensor: RTD, DUT sensors: K thermocouple, Diode, 100Ω RTD, Analog |
Temperature Calibration |
Software calibrated |
Communications Interface |
Ethernet (TCP/IP), USB / Optional IEEE488, RS232 |
Maximum Force Allowable on Thermal Head |
980.6 N (220.4 Lbf) |
DUT Dimensions |
From 2 x 2mm to 50 x 50mm (0.079” x 0.079” to 2.0 x 2.0”) |
Operator Interface |
5.25” color touch-screen, programmable with 0.1°C resolution
Preset temperatures,ramp,soak,cycle.
Ramp rate control, Graphing and data logging, Web server, Offset calibration
Analog temperature inputs |
Thermal Head |
89mm (3.5”) diameter |
Thermal Head Hose |
1.8 meters (70”) long, optional 3.0 meters (118”) long |
Frost Free Thermal Head |
Low flow dry air or nitrogen purge 0.05cfm, -70°C (-94°F) dew point
Controlled automatically |
Physical Dimensions |
610mm D x 445mm W x 445mm H (24”D x 17.5”W x 17.5”H) |
System Weigh |
66 Kg (145lbs.) |
Noise Leve |
55 dBA |
Power Requirements |
115 VAC (±10%), 15 amp, 50/60Hz
208/230 VAC (±10%), 10 amp, 50/60Hz |
Purge Dry Air Supply |
User supplied, regulated, -70°C (-94°F) air or nitrogen, 0.1cfm at 0.2 BAR |
Operating Environment |
Temperature: 5 to 35°C (40 to 95°F), Humidity: 20 to 95% RH |
Compliance |
CE | RoHS | EU 517/2014 | designed to meet UL61010 |